| Model Number |
Implant, 3.5mm x 38mm Anti-Drift Bolt, Fully Threaded |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter states that the anti-drift bolt was removed by the surgeon approximately 3 months after the initial procedure.No pain or symptoms were reported with the screw migration.The joint was fused.The anti-drift bolt was removed, and the patient returned to full activity.
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter states that the anti-drift bolt was removed by the surgeon approximately 3 months after the initial procedure.No pain or symptoms were reported with the screw migration.The joint was fused.The anti-drift bolt was removed, and the patient returned to full activity.Non-conformance was not identified based on the lots that would contribute to this failure mode.The root cause is unknown.Potentially, force from weight-bearing with normal walking motion caused the screw to back-out.Screw back-out is a known failure mode of screws.The following implant were also included in the contruct.Only the anti-drfit bolt was removed.Ref: 7100-lp18-l ln: 501730 qty: 1 product name: dynabunion plate 18mm-left pack-out.Ref: 7118-1818kt-c ln: 300281 qty: 1 product name: dynaforce himaxc implant kit 18x18x18.Ref: 15pl-3516 ln: 400353 qty: 1 product name: motoband cp polyaxial locking screw 3.5mm x 16 mm sterile.Ref: 1500-3522 ln: 400151 qty: 1 product name: motoband cp non-locking screw 3.5mm x 22mm sterile.Ref: 1500-3516 ln: 400342 qty: 1 product name: motoband cp non-locking screw 3.5mm x 16mm sterile.
|
| |
|
Search Alerts/Recalls
|