| Model Number |
1012453-08 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified, mildly tortuous, 80% stenosed lesion in the left anterior descending (lad) artery.A 4.0x8mm nc trek balloon dilatation catheter (bdc) was advanced to the lesion and inflated 3 times to 12, 14 and 14 atmospheres (atm).The balloon fully deflated after the last inflation; however, it was difficult to pull back into the 6f guide catheter after post-dilatation of the stent.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|