• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 6900PTFX
Medical Device Problem Codes Fluid/Blood Leak (1250); Gradient Increase (1270); Difficult to Open or Close (2921); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional customer complaint.The information reported may or may not be related to the edwards device.It was reported via the implant patient registry that this patient with a 6900ptfx 31mm mitral valve, implanted for two years and six months, underwent a valve-in-valve procedure due to unknown reasons.The tmvr was performed with a 9600tfx 29mm.
 
Additional Manufacturer Narrative
The device was not returned to edwards for evaluation as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
 
Event or Problem Description
It was reported via the implant patient registry that this patient with a 6900ptfx 31mm mitral valve, implanted for two (2) years and six (6) months, underwent a valve-in-valve procedure due to severe mitral regurgitation with restricted posterior leaflets and severe mitral stenosis.Patient presented with chronic systolic congestive heart failure.The tmvr was performed with a 9600tfx 29mm.There were no complications.Patient outcome was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
1 edwards way
irvine, CA 92614
9492502731
MDR Report Key15723772
Report Number2015691-2022-08991
Device Sequence Number11495723
Product Code DYE
UDI-Device Identifier00690103163612
UDI-Public(01)00690103163612(17)230619
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P860057/S022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/19/2023
Device Model Number6900PTFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age70 YR
Patient SexFemale
-
-