MAUDE Adverse Event Report: CARDINAL HEALTH UMBILICAL CATHETER SNGL LUMEN 3.5FR; CATHETER, UMBILICAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH UMBILICAL CATHETER SNGL LUMEN 3.5FR; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160333
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported they had an argyle umbilical catheter snap.The catheter snapped near the 24cm marking during removal.The catheter was inserted on (b)(6) 2022 and removed on (b)(6) 2022.It was not replaced.There was no harm to the patient.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record (dhr) review could not be performed because a lot number was not available.However, all dhrs are reviewed to meet regulatory requirements prior to product release.One catheter was received for analysis and investigation.The catheter showed signs of use.It was confirmed that the device showed a rupture near the 23cm mark.The reported issue was confirmed.Based on the information received that the catheter snapped during removal, it was determined that the catheter was in working condition during use and was damaged during removal.There are no further actions required for this complaint because the occurrence is below what is expected and the issue was not confirmed as manufacturing related.This complaint will be used for tracking and trending purposes.
 
Additional Manufacturer Narrative
The device history record (dhr) review could not be performed because a lot number was not available.However, all dhrs are reviewed to meet regulatory requirements prior to product release.One catheter was received for analysis and investigation.The catheter showed signs of use.It was confirmed that the device showed a rupture near the 23cm mark.The reported issue was confirmed.Based on the information available its determined that the catheter was in good condition prior use and the exact root cause of the reported issue could not be determined.This complaint will be used for tracking and trending purposes.
 
Additional Manufacturer Narrative
Additional information: b4: event description was added to due to information received from the customer.
 
Event or Problem Description
Additional information received february 1, 2023: the customer clarified that the catheter did not break when being removed, it was removed because it broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UMBILICAL CATHETER SNGL LUMEN 3.5FR
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
edificio b20 calle #2 zona fra
alajuela
CS  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15724209
Report Number3009211636-2022-00801
Device Sequence Number2872782
Product Code FOS
UDI-Device Identifier20884527005113
UDI-Public20884527005113
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8888160333
Device Catalogue Number8888160333
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer10/31/2022
10/31/2022
10/31/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/13/2022
02/01/2023
03/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-