MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-6-20-3D-SS |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that during the surgery, after the microcatheter was in place, the coil filling was performed.The first coil apb-6-20-3d-ss was chosen to be used.During the delivery of the spring coil, the spring coil was detached in advance when it was not in the ideal state.Later, the micro-guide wire was used for auxiliary adjustment, and the surgery was continued.There was no surgical or medicinal intervention required.The pushwire was not bent or broken.The pushwire will be returned for evaluation.There was no friction or difficulty during delivery.The physician did not reposition the could.The physician did not attempt to detach the coil.The physician did not rotate the delivery pusher during the procedure.Continuous flush was administered during the procedure. no patient symptoms or further complications were reported as a result of this event.The reported device and any accessory devices were prepared as indicated in the ifu. the patient was undergoing surgery for treatment of a saccular, unruptured left side cavernous sinus aneurysm with a max diameter of 6 mm and a 4 mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was normal.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the pushwire was returned for evaluation.There was no implant coil returned with the pushwire.The implant coil appeared to be detached from the pushwire.The shield coil was intact but stretched.The coin was not present against the lumen stop as it was pulling back.The pusher was found broken at the distal break indicator (manual detachment location); retained by the release wire; it appeared that the manual detachment was attempted at this location.The ai and coupler tubing were present and intact; it appeared that the mechanical detachment was not attempted at this location.No bend was observed on the pushwire.All other subassemblies appeared to be normal, and no other anomalies were observed.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations (0.077mm @ 0.063mm; measured 0.089mm @ 0.127mm; measured 0.105mm @ 0.275mm) and found to be within specifications.The inner diameter of the lumen stop, and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00268¿ and found to be within specification.The retainer ring inner diameter (id) was measured to be 0.00460¿ and found to be within specification.Based on the analysis performed, the customer complaint was confirmed as the pushwire was returned with the implant coil already detached from the pushwire.However, the root cause could not be determined.Based on the returned device, there was evidence of manual detachment as the pushwire was found broken at the manual detachment location with the coin pulled back from the lumen stop.Pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.There was no non-conformance to specification identified that that led to the reported issue.Since the implant coil was not returned; any contribution of the implant coil to the reported issues could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|