AMO PUERTO RICO MFG. INC. TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 09/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight and ethnicity: unknown/asked but unavailable.Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.A search of complaints revealed 1 other complaint has been received for this production order number and the issue could not be confirmed as related to manufacturing.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.Attempts have been made to obtain missing information; however, the account did not provide the information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted for an unspecified reason.Through follow-up, it was learned that the lens was explanted from the patient's right eye due to the patient complaining a thick film over vision and made it hard to see.Vision was cloudy, milky, and blurry.The patient could not drive or read.There was no report of any unplanned vitrectomy, incision enlargement, or sutures were required.There was no patient injury.Another johnson & johnson lens, same model zxr00 and higher 24.5 diopter was implanted as a replacement.No other information was provided.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes, section d9: returned to manufacturer on: jan 12, 2023.Device evaluation: the suspect iol was received inside of a specimen cup with an unknown aqueous solution.The lens was cleaned and visual inspection found no issues.No issues were identified with the returned suspect iol.The complaint issue "explant" and "blurry vision" was not identified during product evaluation.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15726127
Report Number3012236936-2022-02750
Device Sequence Number15339089
Product Code POE
UDI-Device Identifier05050474579163
UDI-Public(01)05050474579163(17)260118
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/18/2026
Device Model NumberZXR00
Device Catalogue NumberZXR00U0235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1MTEC30 CARTRIDGE, LOT CK18655
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-