SHIRAKAWA OLYMPUS CO., LTD HALOGEN LAMP Back to Search Results
Model Number MD-151
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "the facility staff attached the halogen lamp (md-151) to the device (otv-si) and lamp did not turn on.The issue occurred at preparation for use.According to the report, the device was replaced as the action taken at the customer site.There is no patient involvement on this reported event.No user injury reported.
 
Additional Manufacturer Narrative
The subject device md-151 ( halogen lamp) was returned for evaluation.Device evaluation has confirmed the customer reported issue.Device evaluation and inspection noted, no abnormalities observed during visual inspection, however, functional testing, noted, the reported problem lamp did not turn on was reproduced , (lamp not lightning).Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Corrected d9, e1 "telephone number" and h8 "reuse" the device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the reported event was likely the result of a deformed socket pin or lamp failure by burn.The root cause is attributed to component failure.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LAMP
Common Device Name
HALOGEN LAMP
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15726739
Report Number3002808148-2022-03787
Device Sequence Number19562996
Product Code FDS
UDI-Device Identifier04953170064319
UDI-Public04953170064319
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K002231
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMD-151
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer06/09/2023
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date06/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OTV-SI
Patient SexUnknown
-
-