• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; HIP INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; HIP INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Codes Separation Failure (2547); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Sweden.Product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a hip procedure, the surgeon surgeon impacted the cup but was not satisfied with the position so the impactor was attached again.After successfully repositioning the cup, the surgeon was not able to get the impactor handle to release from the cup.Potentially the threads were damaged.The impactor was eventually able to be removed from the cup without further incident.There was no harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Common Device Name
HIP INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15729337
Report Number0001825034-2022-02483
Device Sequence Number7186300
Product Code PBI
UDI-Device Identifier00880304532694
UDI-Public(01)00880304532694(11)210818(10)262960
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number262960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexMale
Patient Weight80 KG
-
-