Model Number 802114
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service representative (fsr) installed a new temperature module.Verification/release testing passed successfully.The unit operated to the manufacturer's specifications.
 
Event or Problem Description
It was reported that during priming of the device for a cardiopulmonary bypass (cpb) procedure, the temperature module was flashing yellow and then turned red.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) connected the temperature module to lab use only (luo) testing equipment.The temperature module was ran and tested for 72 hours with no observation of any failures.It was determined that the temperature module met specification.
 
Additional Manufacturer Narrative
The reported complaint was confirmed.Per data log review: the complaint indicated that the issue occurred on (b)(6) 2022, but the log indicated that the issue likely occurred on (b)(6) 2022 (the last time the system was used before (b)(6) 2022).On (b)(6) 2022 the temperature module reported many status changes between in range to under range.This was likely what the reported light emitting diode (led) behavior was.The log confirms the complaint.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key15729412
Report Number1828100-2022-00376
Device Sequence Number11481254
Product Code DTQ
UDI-Device Identifier00886799000748
UDI-Public(01)00886799000748(11)190309
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K172220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number802114
Device Catalogue Number802114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer11/14/2022
12/16/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/05/2022
01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1