| Catalog Number |
07475918190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys igf-1 assay results for 1 patient sample on a cobas e 801 module compared to a siemens immulite.On (b)(6) 2022, the igf-1 result was 89 nmol/l.On (b)(6) 2022, the siemens igf-1 result was 60 nmol/l.The exact date of testing was requested but not provided.On (b)(6) 2022, the sample was repeated on the e801 module and the result was 94 nmol/l.The roche reference range is 15 - 64 nmol/l.The siemens reference range is 23 - 66 nmol/l.No questionable results were reported outside of the laboratory.The e801 module serial number is (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
An aliquot of the patient sample initially reported was provided for investigation.The investigation obtained a roche igf-1 result of 94 nmol/l and a siemens immulite result of 52 nmol/l.The sample was tested using a diasorin analyzer for igf-1.The result was 607.2 ng/ml (79.4 nmol/l).The diasorin reference range for this patient is 83 - 519 ng/ml (10.9 - 67.9 nmol/l).This patient was also reported to be taking norditropin (somatropin).Per product labeling, ".The following special growth disorder drugs were tested.No interference with the assay was found: somatropin - 3.0 mg/l concentration." two additional samples from a second patient were also sent for investigation.Patient 2 is male, 15 years old.The investigation obtained a roche igf-1 result of 46 nmol/l and a siemens immulite result of 29 nmol/l from the first sample.The investigation obtained a roche igf-1 result of 59 nmol/l and a siemens immulite result of 39 nmol/l from the second sample.The roche reference range for patient 2 is 16 - 65 nmol/l.The siemens reference range was requested but not provided.Patient 2's second sample was tested using a diasorin analyzer and the result was 446.0 ng/ml (58.3 nmol/l).The diasorin reference range for patient 2 is 102-520 ng/ml (13.3 - 67.9 nmol/l).The investigation confirmed the recovery of the elecsys igf-1 assay with another competitor method and the who standard.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|