• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD Back to Search Results
Model Number 7171Q/65
Medical Device Problem Codes Fracture (1260); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2022-44203, related manufacturer reference number: 2017865-2022-44205.It was reported that the patient received inappropriate shock from the right ventricular (rv) lead due to noise oversensing.Device interrogation revealed high pacing impedance on the rv lead and premature battery depletion on the implantable cardioverter defibrillator (icd).The physician elected to explant and replace the rv lead and icd.During the procedure the rv lead was noted to be bent and fractured.The left ventricular (lv) lead was dislodged during the procedure and was explanted.The patient condition was stable.
 
Additional Manufacturer Narrative
The reported events were noise oversensing, inappropriate shock, high pacing impedance, bent and fracture.As received, a partial lead was returned in one piece.The reported events of noise oversensing and inappropriate shock were confirmed.Visual inspection of the lead found internal insulation abrasion breaching between ring electrode cable lumen and inner lumen located between the shock coils with abraded ethylene tetrafluoroethylene (etfe) coating of one ring electrode cable and poly tetrafluoroethylene (ptfe) liner of the inner coil.The cause of noise oversensing and inappropriate shock was due to exposure of ring electrode cable and inner coil at the abrasion zone.The reported events of high pacing impedance, bent and fracture were not confirmed.Electrical and x-ray testing did not find any indication of conductor fracture except for cable discontinuities due to procedural damage.No other anomalies found except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR
Common Device Name
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15729702
Report Number2017865-2022-44204
Device Sequence Number16873722
Product Code NVY
UDI-Device Identifier05414734503297
UDI-Public05414734503297
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2013
Device Model Number7171Q/65
Device Catalogue Number7171Q-65
Device Lot Number3781144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer11/15/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
Patient Weight136 KG
-
-