EXACTECH, INC. SMALL SUPERIOR AUGMENT GLENOID PLATE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
SMALL SUPERIOR AUGMENT GLENOID PLATE |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Periprosthetic fracture (4902)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: (b)(4), 300-01-11 - equinoxe, humeral stem primary, press fit 11mm, (b)(4), 315-35-00 - glnd kwire, (b)(4), 320-10-00 - equinoxe reverse tray adapter plate tray +0, (b)(4), 320-15-05 - eq rev locking screw, (b)(4), 320-20-00 - eq reverse torque defining screw kit, (b)(4), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm, (b)(4), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm, (b)(4), 320-20-38 - eq rev compress screw lck cap kit, 4.5 x 38mm, (b)(4), 320-31-40 - glenosphere, 40mm, 320-40-03 humeral liner, 40mm, +2.5 7094722.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that a 71 yo female patient, initial left shoulder implanted on (b)(6) 2022, with no complications, ¿post op x-rays looked perfect¿, underwent a revision procedure on (b)(6) 2022.The x-rays showed that the implant had either rotated superiorly or migrated up.The doctor removed the liner, and then removed the glenosphere and all three screws from the baseplate.He then removed the superior baseplate and found that the top 1/3 superior/posterior had fractured resulting in the baseplate migrating superior.The doctor decided to switch to a tornier baseplate and used a 39 glenosphere.On the humoral side, the stem and +0 tray were well fixed so he trailed with a 40 +2.5 liner but wanted a little more constraint so went with a 40+2.5 liner.The patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of surgical revision due to scapular fracture cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.These devices are used for treatment not diagnosis.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d9, h3, h6 clinical code, component code, type of investigation, investigation conclusion codes.Product has been received by exactech and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|