Model Number 39347-060470
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in a mildly tortuous and mildly calcified cephalic vein.A 6.0mm x 40mm x 75cm athletis pta balloon dilatation catheter was selected for the vascular access interventional therapy case.However, during the first inflation at 20 atmospheres for 10 seconds duration time, the balloon ruptured.The device was able to be removed using the normal method without any issues.Another athletis balloon was used to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
Product code- lit, dqy.
 
Event or Problem Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in a mildly tortuous and mildly calcified cephalic vein.A 6.0mm x 40mm x 75cm athletis pta balloon dilatation catheter was selected for the vascular access interventional therapy case.However, during the first inflation at 20 atmospheres for 10 seconds duration time, the balloon ruptured.The device was able to be removed using the normal method without any issues.Another athletis balloon was used to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
D2b: product code- lit, dqy.E1: initial reporter phone corrected and initial reporter email removed.Device evaluated by manufacturer: an athletis device batch #29625407, 40 atm was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A visual examination identified a longitudinal tear in the distal region of the balloon approximately 6mm in length.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATHLETIS PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15729833
Report Number2124215-2022-44503
Device Sequence Number2995588
Product Code LIT
UDI-Device Identifier08714729973799
UDI-Public08714729973799
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2025
Device Model Number39347-060470
Device Catalogue Number39347-060470
Device Lot Number0029625407
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-