• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS E 411 ANALYZER (RACK SYSTEM); IMMUNOCHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS E 411 ANALYZER (RACK SYSTEM); IMMUNOCHEMISTRY ANALYZER Back to Search Results
Model Number E411 RACK
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys anti-tpo assay on a cobas e 411 analyzer (rack system).The sample initially resulted in a value of 600 iu/ml.The sample was repeated two times, resulting in values of 9.3 iu/ml and 10.7 iu/ml.The questionable result was reported outside of the laboratory.The result of 9.3 iu/ml was deemed correct.The anti-tpo reagent lot number was 62835101 with an expiration date of 30-apr-2022.
 
Additional Manufacturer Narrative
The customer was recommended to centrifuge frozen samples after they are thawed and prior to their testing.The field service engineer replaced the pinch valve tubing and syringe seals.Instrument performance testing passed successfully.The investigation is ongoing.
 
Additional Manufacturer Narrative
The last calibration performed on 11-oct-2022 was normal with no alarms.Quality control data was not provided.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS E 411 ANALYZER (RACK SYSTEM)
Common Device Name
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15731252
Report Number1823260-2022-03485
Device Sequence Number2996278
Product Code JJE
UDI-Device Identifier04015630929702
UDI-Public04015630929702
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K961481
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE411 RACK
Device Catalogue Number04775201001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-