| Model Number |
20E |
| Medical Device Problem Codes |
Failure to Power Up (1476); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported during the event.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported during the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified and duplicated the reported issue.The cause of the reported issue was isolated to the power cable disconnected from the connector designated j17 on the therapy pcba.The cable was reseated to resolve the issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing, and the device was returned to the customer for use.
|
| |
|