• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC REPLANT IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC REPLANT IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 603508
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Per complaint (b)(4), during clinical procedure, implant dropped from the implant driver.
 
Additional Manufacturer Narrative
Implant date and explant date are not applicable since the product was never placed and not removed.Device evaluation results are not available.If the analysis is complete, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Updated b4 for report submission date, g1 for contact information, g3 for awareness date of new information, g6 for report type and sections d9, h1, h2, h3, and h6 to report that the device was not received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPLANT IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
marcos gutierrez
3050 east hillcrest drive
thousand oaks, CA 91362
8184443315
MDR Report Key15731806
Report Number3001617766-2022-05460
Device Sequence Number7542544
Product Code DZE
UDI-Device Identifier10841307103981
UDI-Public10841307103981
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K061319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/17/2022
Device Catalogue Number603508
Device Lot Number95039
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight91 KG
-
-