| Model Number |
CT75STSD-VI |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A sr.Clinical engineer reported an issue with a ct w/7.5fr det cath & vi smartport.During a placement procedure, the distal 29cm of the 66cm catheter broke off in the patient.The catheter was not yet trimmed, when the breakage occurred.Ultimately, the procedure was aborted and the patient underwent an additional procedure to successfully retrieve the catheter segment.The patient did not experience any adverse effects or harm as a result of this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported defective device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of catheter fractured and/or detached cannot be confirmed as no complaint sample was returned. without receiving product for evaluation, we are unable to definitively determine a root cause for this incident. device history record review of the indicated packaging/assembly/catheter lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the instructions for use, which is supplied to the user with this item number, contains the following statements: - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Implantation of attachable catheter: - insert catheter into sheath.Position the distal end of the catheter at the desired location.Peel away sheath while withdrawing it from vessel.Care should be taken not to withdraw catheter as sheath is removed.Catheter position should be confirmed radiographically.Secure catheter in place.- trim proximal end of catheter and advance through subcutaneous tunnel to the port pocket.Attach catheter to the port body.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Returned for evaluation was one (1) smartport and catheter tubing.The customer's complaint description of catheter broken is confirmed.The catheter was returned in two pieces and there was a fracture noted at approximately the 29 cm mark.The catheter met dimensional specifications and there were no manufacturing non-conformances observed.A definitive root cause could not be determined from sample review.The catheter break location is jagged (not clean cut) and located in the middle of the tubing.This would indicate a potential tensile force failure during the catheter tubing tunneling process, however, this cannot be definitively determined.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint. the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use, which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Implantation of attachable catheter: insert catheter into sheath.Position the distal end of the catheter at the desired location.Peel away sheath while withdrawing it from vessel.Care should be taken not to withdraw catheter as sheath is removed.Catheter position should be confirmed radiographically.Secure catheter in place.Trim proximal end of catheter and advance through subcutaneous tunnel to the port pocket.Attach catheter to the port body.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
| |
|