| Model Number |
277KB |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The provided lot code of the cord was gg01 which is from 07/01/2003 which is over the ifu recommendation and is probable cause of wear and tear causing energy to blow off the cable from the connection port of the instrument.The manufacture ifu states the "high frequency cables have a limited life span and recommends replacing hf cables every 3 months if used more than 2-3 procedures weekly".
|
| |
|
Event or Problem Description
|
|
The manufactures reference number (b)(4).During a resection case on (b)(6), the cord 277kb blew off where it attaches to the resectoscope during the case and caught the drapes on fire.Per or manager in follow up reply noted that there was no fire involved, hole in drape.The patient was not harmed and was followed up in recovery.The cord was replaced and the procedure completed.
|
| |
|