| Model Number |
IPN039778 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "after 3 minutes of use, the machine frequently alarmed for helium loss.The issue could not be resolved by adjusting the equipment and the catheter.As a result, the catheter was replaced, and the problem was solved".Additional information states that the secondary catheter was inserted at the same insertion site.The patient condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "after 3 minutes of use, the machine frequently alarmed for helium loss.The issue could not be resolved by adjusting the equipment and the catheter.As a result, the catheter was replaced, and the problem was solved".Additional information states that the secondary catheter was inserted at the same insertion site.The patient condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The serial number on the returned sample is (b)(6).The lot number of the returned sample is 18f22f0034 for the returned sample.Additional information obtained from a teleflex representative indicates the returned sample is the correct iabc for this complaint.Returned for investigation was a 30cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) with the original packaging that does not match the serial number/lot number of the returned sample.The sample was returned in a cardboard box and was loosely packed with the original packaging carton.Returned with the sample was the data scope inflation driveline tubing; some dried blood was noted inside the tubing.Upon return, the one-way valve was tethered to the iabc short driveline tubing.The bladder was fully unwrapped.The iabc central lumen within the flex-tip assembly area was noted damaged; the wire round was unraveled/damaged, and the polyimide (part of the flex tip assembly) was noted no longer attached and separated from the inner cannula (iabc central lumen) at approximately 5.9cm from the iabc distal tip.Dried blood was noted on the exterior surfaces of the returned sample.Dried blood was noted within the bladder/helium pathway.No sheath was returned with the iabc.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times in accordance with quality system document.An attempt to aspirate and flush the iabc central lumen using a 60cc lab-inventory syringe was unable to be successfully completed.Upon aspiration, water was unable to be consistently pulled into the syringe.The attempt to flush resulted in bladder inflation.The results are consistent with the previously noted damaged central lumen.The iabc was leak tested in accordance with testing methods from manufacturing procedure.A leak was immediately detected from the iabc distal tip and iabc luer end.The leak from the iabc distal tip and iabc luer end are consistent with the previously confirmed damaged/separated central lumen.The iabc was leak tested again with the iabc distal tip and iabc luer end blocked off; no other leaks were detected.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.The guidewire met resistance, and the guidewire could not advance at approximately 0.2cm from the iabc distal tip.No blood or debris was noted.The guidewire was front loaded through the iabc luer.The guidewire exited the central lumen and entered the bladder at the location of the inner cannula separation.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for iab "alarmed for helium loss" is confirmed.During the investigation, the iabc central lumen was noted damaged and found no long-er attached to the central lumen flex tip assembly.The damaged central lumen can result in blood entering the helium pathway and helium loss alarms.The iabc bladder was fully intact with no damage or abnormalities noted.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the damaged central lumen.The probable root cause of the complaint is undetermined.A corrective and preventive action (capa) has been initiated to address the issue.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "after 3 minutes of use, the machine frequently alarmed for helium loss.The issue could not be resolved by adjusting the equipment and the catheter.As a result, the catheter was replaced, and the problem was solved".Additional information states that the secondary catheter was inserted at the same insertion site.The patient condition is reported as "fine".
|
| |
|