| Catalog Number |
03P89-24 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded a suspected discrepant pt/inr result on a 75 year old male patient with deep vein thrombosis.There was no additional patient information available, at the time of this report.Return product is not available for investigation.Method: i-stat, date:(b)(6) 2022, result: >4.5.Method: acl top 750, date: (b)(6) 2022, results: 3.8.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 21-nov-2022.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Ak (product complaint level 2 and level 3 investigation procedure).Returned cartridge testing was unable to reproduce the customer's observation of unexpected results.Although the sample size was not sufficient to make a pass/fail determination for rate of results outside allowable error (ea), returned cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev ak.(product complaint level 2 and level 3 investigation procedure) with respect to rate of suppressed results.No deficiency has been identified.
|
| |
|
Search Alerts/Recalls
|