| Model Number |
37370DL |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.
|
| |
|
Event or Problem Description
|
|
The event occurred in china: it was reported that the hook was broken during surgery.The broken part fell into the patient's body but could successfully be retrieved.No harm to patient, user or third reported.Internal karl storz reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The article 37370dl was probably subjected to too much force, which caused the hook to break at the distal end.Since the rest of the hook is still in the instrument and the weld is not damaged, a possible cause is that the hook was overloaded, causing it to bend and beyond which led to the breakage.By reviewing the similar complaint there is no indication for an increase/ trend of the reported issue.Internal karl storz reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is to correct the date in section b4 to reflect the accurate date in which the first supplemental report was submitted on january 5, 2023.Internal karl storz reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction in section h-1 to maulfunction.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|