• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS OPS FIBER 25CM SPOTLIGHT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS OPS FIBER 25CM SPOTLIGHT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number EVLT25OPS
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference pr(b)(4).
 
Event or Problem Description
An end user reported an issue with an evlt kit, with spotlight ops sheath 25cm.During preparation, the fiber was removed from the packaging and the tip was discovered to be bent/cracked.There was no visible damage to the internal or external packaging.The following procedure was completed with another same device and the patient did not experience any adverse effects, harm, or require medical intervention as a result of this incident.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of fiber shaft fractured near tip was not confirmed since no sample was received for evaluation.  without receiving a fiber sample for evaluation a definitive root cause cannot be determined.Potential root cause is handling damage to fiber during transit and/or receiving at customer facility.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.The device history records for the lots obtained through the ship history report (packaging lots) were reviewed.  the review confirmed that the packaging lots met all material, assembly, and performance specification.Labeling review: the directions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPS FIBER 25CM SPOTLIGHT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15735774
Report Number1319211-2022-00048
Device Sequence Number19560120
Product Code GEX
UDI-Device IdentifierH787EVLT25OPS5
UDI-PublicH787EVLT25OPS5
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K041957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue NumberEVLT25OPS
Device Lot Number5733768
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-