MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SPINE SYSTEM SI POLYAXIAL SCREW 5.5 X 7 X 55MM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
| Model Number |
179712755 |
| Medical Device Problem Codes |
Break (1069); Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional device product codes: kwq, mnh, osh, mni, nkb.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in switzerland as follows: it was reported that during a spinal procedure on (b)(6) 2022, screw loosening of l2 with broken screwhead at right side l2 occurred intraoperatively during an extension of a construct from l2-s1 to th1.There was no surgical delay, and the procedure was completed successfully.No fragments were generated.No further information is available.This report involves one expedium spine system si polyaxial screw 5.5 x 7 x 55mm.This is report 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
Surgeon comment: "the connecting segment degeneration th12/l1 does not seem to have been caused by the broken screw head, so that a revision operation would probably have been necessary independently of this.".
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: lot number added.D9: device returned.H4: manufacture date added.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the head of si polyaxl screw 7 x 55mm, p/n: 179712755, broke off.The broken fragment was returned and remains attached to the polyaxial head.No other problems identified.The reported allegation of functional: loose, can not be assessed due to the condition in which the device was returned.A dimensional inspection was unable to be performed due to post-manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the si polyaxl screw 7 x 55mm, p/n: 179712755 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a.Device history lot.A manufacturing record evaluation was performed for the finished device product code: 179712755.Lot number: 255879.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 18/09/2019.Qty: (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5: updated event description.B7: medical history added.
|
| |
|