| Model Number |
560PBT |
| Medical Device Problem Codes |
Contamination (1120); Degraded (1153)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Sore Throat (2396); Respiratory Tract Infection (2420); Cough (4457)
|
| Date of Event |
10/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging hard time breathing, coughing and nasal/throat irritation or soreness.There was no report of serious patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging hard time breathing, coughing and nasal/throat irritation or soreness.Related to a cpap device's sound abatement foam.There was no report of patient harm or injury.After the final attempt to have the device and components returned for evaluation, customer declined to respond to the gfe related questions and terminated the call.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.Section(s) d4 ans h4 were corrected and section h6 updated in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging hard time breathing, coughing and nasal/throat irritation or soreness.Related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.During the exterior investigation, observed unknown dust/dirt contamination present on the base unit and under ui knob.During the internal investigation of base unit found unknown dust/dirt contamination throughout the device internals.Dust contamination was observed in the blower and on the impeller.Sound abatement foam did not appear to have degraded and returned to its original shape when compressed.Evidence of sound abatement foam degradation/breakdown was not observed in this device.Secondary findings: the unknown dust/dirt contamination and fibers found on the blower casing/impeller and blower box was inconsistent with degraded sound abatement foam contamination.The presence of contamination throughout the airpath suggests a source external to the device.The device's downloaded event log was reviewed by the manufacturer and found no errors.The manufacturer concludes unable to directly address the symptoms outlined in the complaint.Confirmed that there was no evidence of sound abatement foam degradation/breakdown observed in the base unit.Confirmed the presence of contamination in the airpath, likely from a source external to the device.
|
| |
|
Search Alerts/Recalls
|