MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 6 FR D/L POWERPICC BASIC TRAY W/70 CM GUIDEWIRE; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 6 FR D/L POWERPICC BASIC TRAY W/70 CM GUIDEWIRE; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Defective Component (2292); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regq3926 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that during catheter insertion, the catheter was found difficult to insert.After removal, the wire and catheter were found deformed.Patient status/ intervention: change new wire.
 
Event or Problem Description
It was reported that during catheter insertion, the catheter was found difficult to insert.After removal, the wire and catheter were found deformed.Patient status/ intervention: change new wire.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a deformed guidewire was confirmed but the cause is unknown.A photograph of the 0.018¿ stainless steel guidewire with a gold-plated tungsten tip was returned for evaluation.The distal tungsten tip was deformed with the tip bent and coiled upon itself.The observable features on the submitted photographs did not contain enough information to identify a specific root cause of the reported event.A review of the manufacture quality and inspection records for the specific lot involved found no potentially related issues encountered during the manufacture and assembly of that product lot.Possible contributing factors for a bent wire could include insertion/removal against resistance or insertion through a tortuous path.H3 other text: evaluation findings are in section h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 FR D/L POWERPICC BASIC TRAY W/70 CM GUIDEWIRE
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15736122
Report Number3006260740-2022-04288
Device Sequence Number14182522
Product Code LJS
UDI-Device Identifier00801741028014
UDI-Public(01)00801741028014
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K050931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number3276155
Device Lot NumberREGQ3926
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-