| Model Number |
4054-9 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported intermittent fault.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.Device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: other, other text: the returned sensor was evaluated.The sensor failed continuity testing and was observed with an open in the cable on the detector circuit at the bend relief area of the sensor end.The sensor failed functional testing and an error message was displayed on the lab monitor.
|
| |
|
Event or Problem Description
|
|
The customer reported intermittent fault.No patient impact or consequences were reported.
|
| |
|