Model Number IPN048872
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The information states that the clinician tested the cuff prior to intubation, and it inflated normally.Once they intubated the patient, the cuff deflated.The clinician had to add 60-70mmhg to inflate the cuff which could potentially harm the patient.The customer confirmed "the patient was doing perfectly fine, and the device did not harm the patient in any way".The patient condition is reported as "fine".
 
Additional Manufacturer Narrative
Qn#qn#(b)(4).Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
The information states that the clinician tested the cuff prior to intubation, and it inflated normally.Once they intubated the patient, the cuff deflated.The clinician had to add 60-70mmhg to inflate the cuff which could potentially harm the patient.The customer confirmed "the patient was doing perfectly fine, and the device did not harm the patient in any way".The patient condition is reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE,CF,7.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15736734
Report Number3003898360-2022-00707
Device Sequence Number6321535
Product Code BTR
UDI-Device Identifier14026704616517
UDI-Public14026704616517
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN048872
Device Catalogue Number5-10114
Device Lot Number73B2200798
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-