| Model Number |
78-50040 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during he surgery, the peek implant did not fit as expected and the surgeon had to alter the implant in order to get it to fit.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed as an intra-operative photo was provided.The bilateral peek implant was split into two pieces in order to be implanted.This caused a surgical delay of approx.25 minutes.The implant was designed as requested by the customer and according to all quality instructions.Further, the dhr rereview showed that the implant was manufactured to specification.
|
| |
|
Event or Problem Description
|
|
It was reported that during he surgery, the peek implant did not fit as expected and the surgeon had to alter the implant in order to get it to fit.
|
| |
|