• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number 2C4040
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address 1: (b)(4).Parque industrial should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that particulate matter (pm) was observed within the tubing of a cysto/bladder irrigation set.The pm was further described as debris inside the tubing that appeared as a bug cooked during sterilization.This issue was identified before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was done which observed contamination-pm (particulate matter) outside the fluid path, between the walls of the tubing.A laboratory analysis was performed which revealed the particulate as, black embedded degraded pieces of burned plastic resin in the tubing.The reported condition of pm was verified.The cause of the condition was due to a manufacturing issue during the extrusion process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOSCOPY IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15737629
MDR Text Key306518550
Report Number1416980-2022-06028
Device Sequence Number1
Product Code LJH
UDI-Device Identifier00085412001050
UDI-Public(01)00085412001050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4040
Device Lot NumberR22D02130
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received11/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-