| Model Number |
100/870/080CZ |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient's tracheostomy cuff was deflating within 15mins.Re-inflation of the cuff was not effective for the post decuffing regime.Early intervention was required.The patient was fasted for an urgent tracheostomy change.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.One used decontaminated sample was returned for investigation.The inflation test was repeated on the received sample.It was found that it is not possible to inflate sample because it leaks so badly.Under closer visual inspection we noticed tear in cuff which caused reported leakage.Each cuff shall be tested by customer prior use as per instruction for use: "the integrity of the cuff and inflation system should be checked prior to insertion".Due to fact that cuff leak was observed after placement it is the most probable that reported failure occurred during tracheostomy procedure due to contact with sharp edge which is in conflict with instruction for use: "guard against cuff damage by avoiding contact with sharp edges".The cause of the reported problem was traced to the user manipulation of the device.No lot number was provided; therefore, dhr (device history review) could not be performed.
|
| |
|
Search Alerts/Recalls
|