Model Number AX082020003
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the evolution transfer carriage and found that the locking pins required adjustment.This allowed the wheels of the transfer carriage to become unlocked resulting in the reported event.The evolution transfer carriage is not under steris service agreement for maintenance activities.The user facility is responsible for all maintenance activities.The technician adjusted the locking pins, tested the unit, confirmed it to be operating according to specification, and returned it to service.The technician counseled facility personnel on the proper use and operation of the transfer carriage.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that their evolution transfer carriage fell to the floor while an employee was unloading the sterilizer.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION TRANSFER CARRIAGE
Common Device Name
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15738829
Report Number3005899764-2022-00054
Device Sequence Number6364934
Product Code FLE
UDI-Device Identifier00724995150358
UDI-Public00724995150358
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAX082020003
Device Catalogue NumberAX082020003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Initial Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-