Model Number M0035421540
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The device was returned and the lot number was confirmed with the packaging returned with the device.During visual inspection the main coil was found to be broken/fractured off the coil delivery wire and not returned.Functional testing was unable to perform as the coil was found to be broken/fractured.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event could not be confirmed; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was analyzed and the main coil was found to be broken off the coil delivery wire and not returned.The coil delivery wire was found to be intact.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefore, an assignable cause of procedural factors will be assigned to the reported event of coil in catheter friction and to the analyzed defect of the main coil broken/fractured during use.
 
Event or Problem Description
The coil (subject device) was returned for analysis and was examined.During device investigation and analysis, it was discovered that the coil (subject device) main coil was found to be broken/fractured during use.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1.5MM X 4CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15738882
Report Number3008881809-2022-00549
Device Sequence Number11502916
Product Code HCG
UDI-Device Identifier04546540697998
UDI-Public04546540697998
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2025
Device Model NumberM0035421540
Device Catalogue NumberM0035421540
Device Lot Number23601097
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2022
Initial Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
Patient Weight58 KG
Patient RaceAsian
-
-