| Model Number |
A22201C |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.This event has been reported by the importer on mdr# 2429304 - 2022 - 00107.
|
| |
|
Event or Problem Description
|
|
Olympus received a voluntary mw reporting an unspecified procedure during which a resection electrode loop 24f broke inside the patient.Olympus is unable to request additional details regarding the patient or reported event, the reporter requested identity and contact information remain confidential.The device will not be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the investigation, the cause of the reported event is likely due to wear and tear of the device.Olympus will continue to monitor field performance for this device.
|
| |
|