MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP; ENDOSCOPES, AC-POWERED AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP; ENDOSCOPES, AC-POWERED AND ACCESSORIES Back to Search Results
Model Number A22201C
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.This event has been reported by the importer on mdr# 2429304 - 2022 - 00107.
 
Event or Problem Description
Olympus received a voluntary mw reporting an unspecified procedure during which a resection electrode loop 24f broke inside the patient.Olympus is unable to request additional details regarding the patient or reported event, the reporter requested identity and contact information remain confidential.The device will not be returned for evaluation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the investigation, the cause of the reported event is likely due to wear and tear of the device.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP
Common Device Name
ENDOSCOPES, AC-POWERED AND ACCESSORIES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22014 5
GM  220145
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key15739610
Report Number9610773-2022-00516
Device Sequence Number2868480
Product Code GCP
UDI-Device Identifier14042761036648
UDI-Public14042761036648
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2023
Device Model NumberA22201C
Device Catalogue NumberA22201C
Device Lot Number1000033994
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer12/23/2022
Initial Report FDA Received Date11/04/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-