Model Number TFNT00
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that the lens was faulty.Additional information was requested.
 
Additional Manufacturer Narrative
The lens was returned, in a non-company peel pouch.Solution was dried on the lens.One haptic was broken/torn from the haptic/optic junction area (not returned).The optic was torn/split/cracked from the haptic/optic junction area towards the center of the optic.Associated products were not provided.It is unknown, if qualified products were used.The root cause could not be determined, for the reported complaint of ¿faulty¿.The specific lens issue was not specified.Damage to the lens was observed.While we are unable to determine, the root cause of the lens damage, our observation reasonably suggests that it is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely, related to customer handling.Company foldable iols are qualified, for use with an company qualified delivery system (handpiece and cartridge), and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications, during the implantation process.Follow up attempts were made.No further information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15740863
Report Number1119421-2022-02362
Device Sequence Number12353336
Product Code MFK
UDI-Device Identifier00380652357838
UDI-Public00380652357838
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFNT00
Device Catalogue NumberTFNT00.250
Device Lot Number15347058
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer03/07/2023
Initial Report FDA Received Date11/05/2022
Supplement Report FDA Received Date04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1