H3, h6.The device was not returned for evaluation.However, the photographs were reviewed, and revealed that the device was fractured.Also, the clinical/medical investigation concluded that, the provided x-ray and photos were reviewed and confirm the reported emperion stem implant breakage at the stem/neck junction.However, as of the date of this medical investigation, supporting clinical documentation has not been provided.Therefore, there were no clinical factors found which would have contributed to the reported event.The patient impact beyond the revision cannot be determined with the limited information provided.Device batch number was not provided; thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed in adverse events that failure to observe the warnings and precautions, trauma, strenuous activity, implant alignment, patient non-compliance and/or involuntary muscular disorders increase the risk of cracking or fracture of implant components, which may lead to revision surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with inspection procedure, the final inspection includes the verification of part configuration per print, it should also verify that the part is free of damaged areas.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include traumatic injury and/or excessive forces.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|