• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was experiencing blurry vision and glare around point sources of light in their left eye (os) resulting in the need to explant the intraocular lens (iol).The patient's pre-operation best corrected visual acuity (bcva) is 20/50.The patient's post-operation uncorrected visual acuity (ucva) is 20/50 and bcva 20/40.The targeted post-operation refraction was 20/20 both eye.No incision enlargement, vitrectomy, or sutures were required.The replacement iol was a different johnson and johnson lens (model diu150,18.5 diopter).There was no reported patient injury.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site, as it is still implanted; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.The search in complaint system revealed 1 other complaint has been received for this po number and it was voided as it was a duplicate serial number.No escalation was required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes section d9: returned to manufacturer on: 11/21/2022 section h3: device evaluated by manufacturer¿ yes device evaluation: the product was returned to the manufacturing site for evaluation.The complaint lens was received inside of a specimen cup.Visual inspection under magnification revealed that the complaint lens was received cut in half.The lens was cleaned and, no issues that could cause or contribute to the complaint issues could be identified.Based on the return condition of the lens, no further product evaluation could be performed.As a result of the investigation there is no indication of a product deficiency or product malfunction.Correction: upon further review of the file, it was noted that the following information (in section h10-narrative field, device evaluation) should be corrected as it is incorrect in initial report, mdr # 3012236936-2022-02745: it was noted that the device was not returned at the manufacturing site, as it was still implanted.However, this is incorrect as the device was already explanted.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: upon further review of the file, it was noted that the following information was inadvertently not included in the initial report (mdr # 3012236936-2022-02745): after the explant, the patient is reported to have good post-operative appearance with visual acuity (va) of 20/30.Additionally, there are no planned interventions for the patient.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15741063
Report Number3012236936-2022-02745
Device Sequence Number6317308
Product Code MFK
UDI-Device Identifier05050474751248
UDI-Public(01)05050474751248(17)241123
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/23/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer11/08/2022
12/16/2022
Initial Report FDA Received Date11/05/2022
Supplement Report FDA Received Date12/02/2022
12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight76 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-