• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7FR X 20CM; CATHETER,INTRAVASCULAR,THERAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7FR X 20CM; CATHETER,INTRAVASCULAR,THERAP Back to Search Results
Model Number IPN916439
Medical Device Problem Codes Unraveled Material (1664); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "after visual verification of the integrity of the device they proceeded to puncture the right internal jugular vein.With ra.Ce.Va, the non-exit of the guide from the needle has been detected.At the attempt to remove the guide, resistance is felt, then the needle is carefully retracted together with the guide from the vessel, and compression is performed.Analysis of the device shows the guide wedged in the needle.Then it comes off the needle and you notice the fraying / unwinding of the thread." no patient harm reported.Another catheter was used and inserted in the left jugular vein.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that "after visual verification of the integrity of the device they proceeded to puncture the right internal jugular vein.With ra.Ce.Va, the non-exit of the guide from the needle has been detected.At the attempt to remove the guide, resistance is felt, then the needle is carefully retracted together with the guide from the vessel, and compression is performed.Analysis of the device shows the guide wedged in the needle.Then it comes off the needle and you notice the fraying / unwinding of the thread." no patient harm reported.Another catheter was used and inserted in the left jugular vein.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 3-LUMEN 7FR X 20CM
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15741173
Report Number3006425876-2022-00939
Device Sequence Number2868525
Product Code FOZ
UDI-Device Identifier10801902165913
UDI-Public10801902165913
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date02/29/2024
Device Model NumberIPN916439
Device Catalogue NumberCV-25703-E
Device Lot Number71F22C2073
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date11/06/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-