The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed due to bending of the light guide bundle.In addition, rattling forceps cap, insufficient bend angle, liquid leakage, yellowing observation, corrosion due to water leakage, grip not watertight, watertightness not maintained, video cable leak, and video connector case discoloration were observed.Lastly, scratches were found on multiple device components.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to stress or user¿s handling.Olympus will continue to monitor field performance for this device.
|