• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 13MM; DENTAL IMPLANT Back to Search Results
Model Number BOST613
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
Date of Event 08/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed due to infection at the implant site.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number: (b)(4).Weight unknown / not provided.Infection: a summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15741409
Report Number0001038806-2022-01684
Device Sequence Number2859453
Product Code DZE
UDI-Device Identifier00844868029724
UDI-Public(01)00844868029724(17)260128(10)2021010527
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/28/2026
Device Model NumberBOST613
Device Catalogue NumberBOST613
Device Lot Number2021010527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Initial Report FDA Received Date11/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-