MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Obstruction of Flow (2423); Infusion or Flow Problem (2964); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Pain (1994)
|
Event Date 10/31/2022 |
Event Type
Injury
|
Event Description
|
Information was received from healthcare provider (hcp) ) via manufacture representative regarding patient receiving dilaudid (hydro morphone) (4.5 mg/24hr/ 25 mg/ml) via implantable pump.It was reported that the patient was not getting pain relief.The environment/external patient factor was when the device was replaced on 2022-04.Attempted catheter access port (cap) and could not aspirated and could not aspirate during surgery.The healthcare provider (hcp) made medical judgment to replace everything as diagnosis could not be determined.The patient was alive and no injury.The weight and medical history were asked but unknown.The patient was alive and no injury.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8781, serial# (b)(4), implanted: (b)(6) 2016, explanted: (b)(6) 2022, product type: catheter.Product id: 8784, serial# (b)(4), implanted: (b)(6) 2016, explanted: (b)(6) 2022, product type: catheter.Other relevant device(s) are: product id: 8781, serial/lot #: (b)(4), ubd: 18-dec-2017, udi#: (b)(4); product id: 8784, serial/lot #: (b)(4), ubd: 14-jan-2018, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a healthcare provider (hcp) via a company representative who reported that the physician suspected a kink in the catheter was the cause of the issue, but the cause was unknown.
|
|
Event Description
|
Additional information received from the company representative stated the the patient was receiving dilaudid (25.0 mg/ml and 4.505 mg/day ) and bupivacaine (15.0 mg/mland 2.703 mg/day).
|
|
Manufacturer Narrative
|
Continuation of d10: product id 8781 lot# serial# (b)(6) implanted: (b)(6) 2016 explanted: (b)(6) 2022 product type catheter product id 8784 lot# serial# (b)(6) implanted: (b)(6) explanted: (b)(6) 2022 product type catheter medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The pump and the catheter was returned and analyzed.Analysis found no anomaly on the pump, no significant anomaly on the catheter anchor and a catheter body kinked.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|