Model Number U42002H10RX
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 08/2023).
 
Event or Problem Description
It was reported that during an angioplasty procedure in left radial artery, the pta balloon allegedly detached.It was further reported that the balloon allegedly difficult to remove.Reportedly, fragment was removed successfully using snare.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse rx in two segments was received for evaluation.Additionally, an unknown sheath with snare inserted at the distal end was also returned.Visual evaluation showed segment 1 consisted of the hub and catheter portion that had a complete circumferential break.Segment 2 consisted of the remaining portion of catheter and balloon still loaded onto a guidewire.No functional testing was performed due to the nature of the complaint.Therefore, the investigation is confirmed for the reported catheter detachment as the device was returned detached in two segments.However, the investigation is inconclusive for the reported sheath removal difficulty as functional testing could not be performed due to the nature of the complaint.A definitive root cause for the alleged catheter detachment and sheath removal difficulty could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2023), g3 h11: b5, e1, e4, g2, h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in left radial artery, the pta balloon was allegedly difficult to remove through the sheath.It was further reported that the balloon allegedly sheered in half.Reportedly, the detached parts were able to be removed completely from the body using a snare.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE RX
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15741882
Report Number2020394-2022-00868
Device Sequence Number7015406
Product Code LIT
UDI-Device Identifier00801741056574
UDI-Public(01)00801741056574
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K131199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU42002H10RX
Device Catalogue NumberU42002H10RX
Device Lot NumberCMEW0331
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date12/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-