| Model Number |
CCWET3 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non health care professional reported that after intraocular lens implantation, the patient experienced monocular diplopia.The lens was explanted and replaced with a monofocal lens within 2 months of the initial procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information has been requested and received stating that the iol was exchanged for the another lens model but three and half diopter less power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.The manufacturer internal reference number is: (b)(4).
|
| |
|