The customer reported via phone call that they were hospitalized due to high blood glucose on (b)(6) 2022 with blood glucose level of 600 mg/dl.Customer experienced symptoms such as feeling sick, headache, tired and weak at the time of hospitalization.Customer's current blood glucose value was 133 mg/dl.Customer used insulin pump system within 48 hours of reported event.Auto mode feature was active at the time of event.Customer treated for high blood glucose with manual injection.Customer tested for ketone, the result was 15.Customer alleged insulin pump was under delivering.The insulin pump will not be returned for analysis.
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|