| Model Number |
ONYXNG30012UX |
| Medical Device Problem Codes |
Malposition of Device (2616); Device Dislodged or Dislocated (2923); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use one onyx frontier coronary drug eluting stent to treat a moderately tortuous, moderately calcified lesion in the proximal left anterior descending (lad) artery with 90% stenosis.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was not encountered when advancing the device.It was reported that stent dislodgement occurred during delivery to the lesion.It was detailed that the stent was delivered beyond the lesion site and as the stent was pulled into place it appeared that the proximal edge of the device got caught on the strut of the previously implanted stent.The previously deployed stent was distal to the identified lesion.The stent distal to the identified lesion had an area in the proximal segment which appeared to be under-deployed.The dislodged stent was successfully removed with a snare.No further patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the device was received for analysis.The stent was not present on the balloon on return for analysis.A kink was evident on the distal shaft.The balloon folds remained intact.Crimp impressions were visible on the exposed balloon surface.The stent returned on a snare.Deformation was evident to the dislodged stent.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|