Catalog Number 04469658190
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer (fse) cleaned the water tank and water conductivity was acceptable.The fse noted that the sample in question was run in a 13 mm sample tube with no rack adapter.The customer¿s qc data showed very high results.The customer used an x50 dilution for the initial test on the c502 module.Product labeling states: "determine samples having higher concentrations via the rerun function.Dilution of samples via the rerun function is a 1:11 dilution.Results from samples diluted using the rerun function are automatically multiplied by a factor of 11." the investigation is ongoing.
 
Event or Problem Description
The initial reporter complained of discrepant results for 1 patient urine sample tested for albt2 tina-quant albumin gen.2 (albt2) on a cobas 8000 c 502 module.The customer first tested the sample manually using a test strip and the result was "positive." the customer then ran the sample on the c502 module using an x50 dilution with a result of 14.0 mg/l.This result was reported outside of the laboratory.The repeat result using an x20 dilution was 1411.6 mg/l.On (b)(6) 2022 the sample was repeated using an x10 dilution and the result was 1410.3 mg/l.The c502 module serial number was (b)(4).
 
Additional Manufacturer Narrative
The customer used an x50 dilution for the initial test on the c502 module.A 1:50 dilution is not reasonable for the initial test and could have led to the questionable low result.Product labeling states: "determine samples having higher concentrations via the rerun function.Dilution of samples via the rerun function is a 1:11 dilution.Results from samples diluted using the rerun function are automatically multiplied by a factor of 11." in this case, a valid result could have been received by using the decreased mode.The result of 1411.6 mg/l was believed to be correct.The investigation determined the event was consistent with a handling error (dilution factor) at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALBT2 TINA-QUANT ALBUMIN GEN.2
Common Device Name
ALBUMIN IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15742544
Report Number1823260-2022-03508
Device Sequence Number5661119
Product Code DCF
UDI-Device Identifier04015630920587
UDI-Public04015630920587
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
K101203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number04469658190
Device Lot Number65268801
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-