| Model Number |
C311 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a questionable creatinine jaffe, gen 2 result from the cobas 4000 c311 system.The initial result was 4.43 mg/dl with a data flag and was reported outside of the laboratory to the doctor.As the result did not coincide with the patient's medical history, the doctor requested repeat testing.On 13-oct-2022, the same sample was retested on the same analyzer with a result of 2.09 mg/dl and on a cobas c501 analyzer with a result of 1.80 mg/dl.The repeat result was believed correct.The reagent lot number was 63597001 with an expiration date of 31-jan-2024.
|
| |
|
Additional Manufacturer Narrative
|
|
A general reagent problem could be ruled out because calibration and quality control were acceptable.It was found that the customer is not cleaning the probes on a daily basis as recommended.Therefore a carry-over event can not be excluded.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|