ROCHE DIAGNOSTICS COBAS C 311 ANALYZER; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C311
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of a questionable creatinine jaffe, gen 2 result from the cobas 4000 c311 system.The initial result was 4.43 mg/dl with a data flag and was reported outside of the laboratory to the doctor.As the result did not coincide with the patient's medical history, the doctor requested repeat testing.On 13-oct-2022, the same sample was retested on the same analyzer with a result of 2.09 mg/dl and on a cobas c501 analyzer with a result of 1.80 mg/dl.The repeat result was believed correct.The reagent lot number was 63597001 with an expiration date of 31-jan-2024.
 
Additional Manufacturer Narrative
A general reagent problem could be ruled out because calibration and quality control were acceptable.It was found that the customer is not cleaning the probes on a daily basis as recommended.Therefore a carry-over event can not be excluded.The investigation is ongoing.
 
Additional Manufacturer Narrative
The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS C 311 ANALYZER
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15742555
Report Number1823260-2022-03511
Device Sequence Number20601519
Product Code JJE
UDI-Device Identifier04015630922284
UDI-Public04015630922284
Combination Product (Y/N)N
Initial Reporter CountryUY
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC311
Device Catalogue Number04826876001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-