| Model Number |
Ø9X 20MM BC IF SCRW, VENTED |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee surgery the screw broke at the beginning of the insertion.The broken off part was retrieved from the patient.According to the surgeon, no harm for patient, operator or third party occurred.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Upon visual inspection it was noted that a fracture was present at the tip of the screw and present throughout the body of the screw between the spiral threads.The cause of this event is undetermined.The most probable causes for this include improper bone prep, misaligned insertion, prying leveraging the device during insertion, and/or shallow driver engagement depth.
|
| |
|