| Model Number |
ACROBAT-I POSITIONER |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
| |
|
Event or Problem Description
|
|
The hospital reported that during a coronary artery bypass procedure the acrobat-i positioner ch arm would not tighten all the way enough to hold the heart up.They opened up another one to finish the case with no patient affects.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise # (b)(4).The device was returned to the factory for evaluation on 11/01/2022.An investigation was conducted on 01/02/2023.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.A mechanical evaluation was conducted.The positioner was attached and locked onto a reference retractor blade by sliding the stabilizer base onto the rail and locking the lever.The flexlink arm was able to be adjusted and positioned, and then locked into place by turning the knob clockwise with no physical or visual difficulties, with the locking lever in both the locked and unlocked positions.Based on the returned condition of the device as well as the evaluation results, the reported failure ¿mechanical problem" was not confirmed.The lot # 3000225770 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|