• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNIVERS REVERS TRIAL 36 +3; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERS REVERS TRIAL 36 +3; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number UNIVERS REVERS TRIAL 36 +3
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via sems that a ar-9530s-03 trial humeral insert broke through the holes, not allowing us to insert the liner.This occurred during an unknown case, with no patient effect reported.
 
Additional Manufacturer Narrative
The complaint was confirmed.Visual evaluation of the pictures attached to the complaint of an ar-9530s-03 univers revers trial 36 +3.It is concluded that the symmetry of the circle was damaged.The material around the circle shows nicks and is missing a piece.The most likely cause for the reported failure is user error for applying excessive force during use.Per dfu-0023- at revision 3.I.Cautions.2.To avoid damaging the instruments, do not impact or subject to blunt force any instruments that are designed to be turned or screwed in.When two devices are intended to be threaded together, ensure that they are fully engaged prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS REVERS TRIAL 36 +3
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15743519
Report Number1220246-2022-05712
Device Sequence Number5850499
Product Code LXH
UDI-Device Identifier00888867062146
UDI-Public00888867062146
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUNIVERS REVERS TRIAL 36 +3
Device Catalogue NumberAR-9530S-03
Device Lot Number345139
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date10/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-