MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CONV FB PS TB TRL SZ4; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CONV FB PS TB TRL SZ4; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-00-974
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the surgeon complained that the drill was dull.The impactor handle broke.Red tip broke off.All being sent back.Trial poly and shim cracked.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned.After review of the returned device, the reported condition of the attune conv fb ps tb trl sz4 [254500974./bfa0hea] is confirmed.The surface of the device has cracked, also one of the springs has not returned.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV FB PS TB TRL SZ4
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15743643
Report Number1818910-2022-22193
Device Sequence Number7133433
Product Code HWT
UDI-Device Identifier10603295135517
UDI-Public10603295135517
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2545-00-974
Device Catalogue Number254500974
Device Lot NumberBFA0HEA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Initial Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/26/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
-
-